SOFTWARE ALERT: Midmark ECG RR Option
Dear
PCM Client,
The
following problem has been reported when
using the Midmark ECG RR function; the
standard ECG is not affected.
Affected
PCM versions: All Midmark versions
Patch
Update: Will require Midmark
8.4 update, available in approximately
1 week
Description
of the issue: When using
the "Start RR" function, the
ECG RR files are not being saved in the
correct ECG MM Chart Tab, therefore these
files are not visible in the patient's
chart. Note: These files are not lost;
they have to be copied/re-imported in
the correct directory.

Solution/Recommendation:
Please do
not use the Start RR function until this
issue is resolved. If you are currently
using this function and have experienced
this issue, please contact support@primeclinical.com and we will copy/re-import
the files into the correct location.
We
apologize for the inconvenience.
Sincerely,
Prime Clinical Systems, Inc.
PCM
Support
|